Tools Required - Keysight Technologies 8970B Quick Start Manual

Noise figure meter security and memory clearing

Advertisement

8970B Noise Figure Meter Security and Memory Clearing
Introduction
8970B memory clearing process requires removal of the processor battery for 15 minutes or longer.
Below are step-by-step instructions for removing this battery. Make sure you use proper ESD
pre-cautions.

Tools Required

— Torx T15 driver
— #2 Phillips screwdriver
Procedure
1. Unplug the 8970B from AC power.
2. Place the instrument upside-down on a flat surface.
3. Remove the bottom cover using Torx T15 driver and #2 Phillips screwdriver.
4. Lift the processor assembly by loosening the two screws and hinge the board away as shown
below:
Figure 1
Battery Location
8970B Noise Figure Meter Security and Memory Clearing
8970B Noise Figure Meter Security and Memory Clearing