Regulatory Information; Safety Compliance; Emc Compliance - Keysight 8990B Installation Manual

Peak power analyzer
Hide thumbs Also See for 8990B:
Table of Contents

Advertisement

Regulatory Information

The 8990B complies with the following safety and Electromagnetic Compatibility
(EMC) compliances:

Safety compliance

The 8990B conforms to the requirements of the following safety standards:
– IEC 61010- 1:2001/EN 61010- 1:2001
– CAN/CSA- C22.2 No. 61010- 1- 04
– ANSI/UL std No. 61010- 1:2004

EMC compliance

The 8990B complies with the essential requirements of the following applicable
European (EC) Directives, and carries the CE marking accordingly to the Low
Voltage Directive (2006/95/EC) and EMC Directive (2004/108/EC).
EMC tests conform to the IEC 61326- 1:2005/EN 61326- 1:2006 and CISPR
11:2003/EN 55011:2007 (Group 1, Class A). In order to preserve the EMC
performance of the 8990B, any cable which becomes worn or damaged must be
replaced with the same type and specification.
The 8990B also meets the following EMC standards:
– Canada: ICES/NMB- 001: Issue 4, June 2006
– Australia/New Zealand: AS/NZS CISPR11:2004
Degradation of some instrument specifications can occur in the presence of
ambient EM fields and noise that are coupled to the power line or I/O cables of the
8990B. The 8990B will self-recover and operate to all specifications when the
source of ambient EM fields and noise are removed or when the 8990B is
protected from the ambient EM fields or when the 8990B cabling is shielded from
the ambient EM noise.