Positioning The 8990B - Keysight 8990B Installation Manual

Peak power analyzer
Hide thumbs Also See for 8990B:
Table of Contents

Advertisement

1
Getting Started

Positioning the 8990B

Position the 8990B where it will have sufficient clearance for airflow around the
top, rear, and sides (refer to the following figure). Also, only place the 8990B on a
hard surface to prevent blocking the airflow underneath the 8990B (for example, a
piece of paper or carpet can block the fans and cause the 8990B to overheat).
Minimum side clearance:
50 mm
Figure 1-1
26
Minimum top clearance: 0 mm
8990B
(Front view)
Minimum bottom clearance: No intrusion into the
space under the 8990B as defined by the feet. The
feet must rest on a hard surface.
Rear panel
Minimum rear clearance: 0 mm
8990B
(Top view)
The airflow requirement is 300 cfm
Positioning the 8990B for proper airflow
Minimum side clearance:
50 mm
Keysight 8990B Installation Guide