Rigol DS2000E Series User Manual page 272

Digital oscilloscope
Hide thumbs Also See for DS2000E Series:
Table of Contents

Advertisement

RIGOL
Time Base Mode
Number of X-Ys
Waveform Capture Rate
Vertical
Bandwidth
(-3 dB) (50 Ω)
Single-shot Bandwidth (50 Ω)
Vertical Resolution
[3]
Vertical Scale
Offset Range
[1]
Bandwidth Limit
Low Frequency Response
(AC Coupling,
-3 dB)
Calculated Rise Time
DC Gain Accuracy
DC Offset Accuracy
Channel-to-Channel Isolation
16-2
Y-T, X-Y, Roll
1 path
[2]
50,000 wfms/s (dots display)
DS2102E: DC to 100 MHz
DS2202E: DC to 200 MHz
DS2102E: DC to 100 MHz
DS2202E: DC to 200 MHz
8 bits
When the input impedance is 50 Ω: 500 μV/div
to 1 V/div
When the input impedance is 1 MΩ: 500 μV/div
to 10 V/div
When the input impedance is 50 Ω:
500 μV /div to 50 mV/div: ±2 V
51 mV/div to 200 mV/div: ±10 V
205 mV/div to 1 V/div: ±12 V
When the input impedance is 1 MΩ:
500 μV /div to 50 mV/div: ±2 V
51 mV/div to 200 mV/div: ±10 V
205 mV/div to 2 V/div: ±50 V
2.05 V/div to 10 V/div: ±100 V
DS2102E: 20 MHz
DS2202E: 20 MHz/100 MHz
≤5 Hz (on BNC)
[1]
DS2102E: 3.5 ns
DS2202E: 1.8 ns
[3]
±2% of full scale
±0.1 div±2 mV±1% of offset value
DC to maximum bandwidth: >40 dB
Chapter 16 Specifications
DS2000E User's Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds2202eDs2102e

Table of Contents