Process signals source
Table of Contents

Advertisement

Step 1: Using the 〔V〕 key to select DC voltage source function, select the desired range from 100mV, 1000mV,
and 10V by pressing the〔RANG〕key. The default value and unit of the selected source function and range
shall be displayed in the main districts part of the LCD.
Step 2: Set the output value digit by digit using 〔〕/〔〕keys.
Each pair of 〔〕/〔〕keys corresponds to each digit of the LCD reading. Each press of the 〔〕/〔〕
key increases or decreases the digit. Increasing the digit from 9 or decreasing it from 0 causes the digit to
overflow or underflow, allowing you to set the output value without interruption. Holding down the 〔〕/〔〕
key continuously changes the digit in question. And the value won't change if it is increased or decreased to
the Maxim or Minimum value. Pressing the〔ZERO〕key initializes the output set point to the default value (0).
Step 3: Pressing the 〔ON〕 key causes the indicator on the LCD to change from ''OFF'' to ''ON ''. The calibrator
sources the preset DC voltage between the output terminals.
Step 4: To turn off the output, press the 〔ON〕 key once again. The'' OFF '' appears on the LCD and no signals
sourced between the terminals.
7.3 Sourcing DC Current
Step 1: Using the〔mA〕key to select the desired source function 0-22mA .The default value and unit of the
selected source function shall be displayed in the main districts part of the LCD.
Step 2: Set the output value digit by digit using〔〕/〔〕keys.
Each pair of〔〕/〔〕keys corresponds to each digit of the LCD reading. Each press of the〔〕/〔〕
key increases or decreases the digit. Increasing the digit from 9 or decreasing it from 0 causes the digit to
overflow or underflow, allowing you to set the output value without interruption. Holding down the〔〕/〔〕
key continuously changes the digit in question. And the value won't change if it is increased or decreased to
18