100(current output value mA-4mA)
mA %=
16 mA
Tips:
Press the (T.DISP) key, assistance districts converts the room temperature.
7.3.4 4-20 mA simulate transmitter source
Connect the calibrator and the loop power as listed in Figure 9, and operate in steps shown in sourcing DC
current.
Figure 9 4-20 mA simulate transmitter source
7.4 Sourcing Resistance
Firstly, the calibrator sources a resistance signal by receiving the resistance-measuring current I supplied
from the device being calibrated (such as a resistance meter) and then delivering the voltage V
%