Process signals source
Table of Contents

Advertisement

DCV_100mV
DCV_1000mV
DCV_10V
DCmA_20mA
OHM_400Ω/1mA
OHM_400Ω/0.1mA
OHM_4KΩ/0.1mA
OHM_40KΩ/0.01mA
﹡Adjusting the displayed value same with the reading of the digit meter when the present calibrator is
stabilized.
You can calibrate a desired function and range separately.
You must calibrate all the calibrating points of the selected range together.
When adjusting resistance source, the exciting current is (+) for adjustment point ''0'' and ''F'', and is (-) for
adjustment point "–0" and "–F".
Turn on the meter; press the (START) key while simultaneously holding down the〔ON〕key enters the source
calibration state. LCD shows "CAL-0" symbol on the assistance districts part, the present calibrating point on
the main districts part and the high 5 digits of the responding value and its unit. The digit in the right on the
assistance districts part is the lowest digit of the value.
Tips:
0
/
100mV
0
0
1000mV
0
/
10V
0
/
20mA
400Ω
/
/
400Ω
/
4KΩ
/
40KΩ
/
/
/
/
/
/
/
/
-0Ω
-400Ω
I=±1mA
-0Ω
-400Ω
I=±0.1mA
-0Ω
-4KΩ
I=±0.1mA
-0Ω
-40KΩ
I=±0.01mA