Accuracy is expressed as ± (percentage of set value + percentage of range)
Function
Referenc
Range
e
DC voltage 100mV
-10.000mV
110.000mV
1000mV
-100mV
1100mV
10V
-1.0000V
11.0000V
DC current
20mA
0.000mA
22.000mA
Resistance 400Ω
0.00Ω~400.00Ω
4KΩ
0.0000
4.0000 KΩ
Resoluti
Accuracy
on
1μV
0.02+0.01
10μV
0.02+0.01
0.1mV
0.02+0.01
1μA
0.02+0.02
0.01Ω
0.02+0.02
KΩ ~
0.1Ω
0.05+0.025
Remark
Maximum output current: 0. 5mA
Maximum output current: 2mA
Maximum output current: 5mA
External supply for simulate mA:
5V–28V
Maximum load 1KΩ at 20mA
Excitation current: ± 0.5–3 mA;
if ± 0.1–0.5, add 0.1Ω;
Accuracy does not include lead
resistance;
Excitation current: ±0.05
–0.3mA;
Does not include lead resistance;