as below:
AP.OFF: 10min.
BL.OFF: 10sec.
TMP.U: ℃.
Tips:
Any change of setting to the above-mentioned function, press the 〔ON〕 key to save the value. Any press of
the〔ON〕key saves the nearest setting value.
9 Adjusting Source Functions
Environmental Requirements
Ambient temperature: 23 +2℃
Relative humidity: 35% to 75% RH
Warm-up:
Before using, warm up the calibrator for the period of time specified.
Set the meter into the standard environment for 24 hours, and then turn on the power. Change the set into
non-automatic power-off state and warm it up for one hour.
Power Supply: new alkaline size (AAA) battery type 7 is the best choice for adjustment.
Source Adjustment Operation:
Table 4 Adjustment Points of Source Functions
Range
Adjustment
Point *
0
A
F
Remarks
-0
-F
33