Maintenance
3.
Lift the power supply to expose the ground wire (2) and wiring harness (3).
4.
Remove P2 from J2 on the power supply assembly and the ground wire
connection (2) from the power supply chassis.
5.
Reverse the previous steps to attach a new power supply.
NOTE:

Top Cover

It is often necessary to remove the top cover in order to service other components in
the device. The following sections explain how to remove and replace the top cover.
Removing the Top Cover
Use the following procedure to remove the top cover.
NOTE:
It is not necessary to remove the power supply prior to removing the top cover.
A #10 Torx driver is required for this procedure.
1.
Remove the battery.
Refer to
2.
Turn the unit over so the bottom side is facing up and remove the TORX screw
through the hole on the right rear corner of the unit.
104
Before replacing the screws (1), ensure that the ground wire (2) is routed
through the notch in the plastic and not pinched.
"Replacing the Battery" on page 99
MAC™ 5500/MAC™ 5500 HD
for details.
2046275-017G