Maintenance
10.
11.
Diagnostic Tests
12.
13.
14.
15.
16.
17.
18.
19.
20.
Electrical Safety Checks
21.
1.
HD is displayed only if the following conditions are met: (1) a CAM HD is attached, (2) v10 software is installed, and (3) the HDMD option
is activated.
134
Trolley casters lock successfully?
HD displayed on the ECG screen?
Battery Status Test meets Battery Current expectation?
Display diagnostic test successful?
Speaker test successful?
Keyboard test successful?
Writer diagnostic tests successful?
Read/Write Floppy Diagnostic Test successful? (Accurate test does not have to be
performed)?
External/Internal Modem Test successful?
SD Card Test successful?
Ethernet Module Test successful?
Current leakage and ground continuity test results meet requirements?
Perform electrical safety checks when indicated. All indicated electrical safety
checks require a pass/fail indication for steps performed. Record the measurement
values in your debrief.
MAC™ 5500/MAC™ 5500 HD
1
2046275-017G