2025177-031 Top Cover Gasket

Revision C
Parts List, Drawings, and Replacement: Assembly/Disassembly of FRUs
1. Remove top cover. Remove the gasket on the inside front edge by
pulling it free of the cover.
2.
Clean the stamping area where the gasket was adhered to the cover by wiping
the surface with isopropyl alcohol.
3.
Align the front edge of the gasket along the stamping line on the inside of the
top cover and apply pressure to secure it. Ensure the adhesive of the gasket is
completely sealed all along the stamping line.
CAUTION
The top cover gasket protects the inside of the unit from fluid
spills. Whenever the monitor cover is removed during servicing,
be sure to inspect the gasket, ensuring it is attached securely
across the whole front of the top cover along the stamping line
(just in from the edge). If the adhesive is no longer effective, order
replacement Top Cover Gasket FRU 2025177-031.
250cx Series Maternal/Fetal Monitor
2036947-001
6-31