Language Reference
SYSTem Subsystem
Factory Preset
(no *RST):
The factory default is IBFull. This parameter is
persistent, which means that it retains the setting
previously selected, even through a power cycle.
Front Panel
Access:
none
Serial Port Baud Rate Setup
:SYSTem:COMMunicate:SERial[1]|2|3|4|5|6|7|8[:RECeive]:BAUD
<baud_rate>
:SYSTem:COMMunicate:SERial[1]|2|3|4|5|6|7|8[:RECeive]:BAUD
?
There can potentially be four I/O cards installed (1, 2, 3, or 4). Each
card may have up two ports:
Card one uses port numbers 1 and 2
Card two uses port numbers 3 and 4
Card three uses port numbers 5 and 6
Card four uses port numbers 7 and 8
Factory Preset
(no *RST):
The factory default is 9600. This parameter is
persistent, which means that it retains the setting
previously selected, even through a power cycle.
Range:
Supported baud rates are
110|300|600|1200|2400|4800|9600|19200|38400|
57600|115200
Front Panel
Access:
System, Remote Port
Serial Port Receive Pace Setup
:SYSTem:COMMunicate:SERial[1]|2|3|4|5|6|7|8[:RECeive]:PACE
XON|NONE
:SYSTem:COMMunicate:SERial[1]|2|3|4|5|6|7|8[:RECeive]:PACE
?
Set the receive pace to on or none for an instrument, with the RS-232
interface installed. If no optional serial port number is specified, port 1
is assumed.
There can potentially be four I/O cards installed (1, 2, 3, or 4). Each
card may have up to two ports:
Card one uses port numbers 1 and 2
5-118
Chapter 5