CALCulate Subsystem
Front Panel
Access:
Marker, <active marker>, Start
Band Markers Stop Frequency X Position
:CALCulate:MARKer[1]|2|3|4:X:POSition:STOP <param>
:CALCulate:MARKer[1]|2|3|4:X:POSition:STOP?
Position the right-most marker, the stop frequency of the designated
band-type marker pair, at the specified trace X position. A trace is
composed of 401 points (X positions.) Use :CALCulate:MARKer:MODE
BAND to select band markers.
The query returns the current X-position stop frequency of the
designated marker.
Range:
0 to 400
Default Unit:
X-axis units (Hz or seconds)
Front Panel
Access:
Marker, <active marker>, Stop
Marker X-Axis Readout
:CALCulate:MARKer[1]|2|3|4:X:READout
FREQuency|TIME|ITIMe|PERiod
:CALCulate:MARKer[1]|2|3|4:X:READout?
Selects the units for the x-axis readout of the marker. Available units
are:
Frequency
Time
Inverse of time
Period
Factory Preset
and *RST:
Frequency
Front Panel
Access:
Marker, Readout, Frequency
Marker, Readout, Time
Marker, Readout, Inverse Time
Marker, Readout, Period
Span-Markers Span X Value
:CALCulate:MARKer[1]|2|3|4:X:SPAN <param>
:CALCulate:MARKer[1]|2|3|4:X:SPAN?
Change the frequency span of the designated span-type marker pair to