Download Print this page

Technical Data; Measured Values - E+E Elektronik EE210 User Manual

Humidity and temperature transmitter for demanding climate control applications
Hide thumbs Also See for EE210:

Advertisement

TECHNICAL DATA

(Modification rights reserved)

Measured Values

Relative Humidity (RH)
Sensor
Working range
RH accuracy (incl. hysteresis, non-linearity and repeatability)
Wall & duct version:
-15...40°C (5...104°F)
-15...40°C (5...104°F)
-40...60°C (0...140°F)
Remote probe version
at 20°C (68°F) ±2.5% RH
Temperature (T)
Sensor
T-accuracy
Outputs
Analog output
(RH: 0...100%; T: see ordering guide)
Digital output
General
Power supply
for 0-5 V / 0-10 V / RS485
for 4-20 mA
Current consumption
Voltage output
Current output
Digital interface
Display
Connection
Housing material
Protection class
Cable gland
Probe cable (type C)
Sensor protection
Electromagnetic compatibility
Temperature ranges
Temperature ranges with display
1 USA & Canada: class 2 supply required, max. supply voltage 30V
E+E Sensor HCT01-00D
0...100% RH
≤90% RH
>90% RH
±(1.5 + 0.015*measured value) % RH
Pt1000 (tolerance class B, DIN EN 60751) integrated in HCT01
wall & duct
∆ °C
0-5 V / 0-10 V
4-20 mA (two-wire)
RS485 (BACnet MS/TP or Modbus RTU), max. 32 EE210 in one bus
15 - 35V DC
10V + RL x 20 mA < V+ < 30V DC
DC supply typ. 3.3mA; with display typ. 3.6mA
AC supply typ. 34mA; with display typ. 37mA
DC supply max. 40mA
DC supply typ. 5mA; with display typ. 19mA
AC supply typ. 52mA; with display typ. 118mA
1, 2 or 3 lines, user configurable
Screw terminals, max. 1.5 mm
Polycarbonate, UL94V-0 (with Display UL94HB) approved
IP65 / NEMA 4
M16 x 1.5
PVC, Ø 4.3mm, 4 x 0.25 mm2, Length: 1.5 or 3m (4.9 or 9.8ft)
E+E Coating
EN61326-1 EN61326-2-3
Industrial Environment
Operating: -40...60°C
Storage:
Operating: -20...50°C
Storage:
±(1.3 + 0.003*measured value) % RH
± 2.3% RH
∆ °C
°C
-1 mA < I
< 1 mA
L
R
≤ 500 Ohm
L
or 24V AC ±20%
1)
2
(-40...80°C for remote probe EE210P)
(-40...140°F)
-40...60°C
(-40...140°F)
(-40...80°C for remote probe EE210P)
(-4...122°F)
-20...60°C
(-4...140°F)
remote probe
°C

Advertisement