To Replace The Battery; To Clean The Airflow Sensor Orifice - Carrier VVT Bypass II Installation And Startup Manual

Table of Contents

Advertisement

To replace the battery

If the VVT Bypass II experiences a power outage and the control program stops functioning, replace the
battery.
You need to replace the battery if the voltage measures below 2.9 volts when the controller is not powered.
1
Verify that the VVT Bypass II's power is on.
CAUTION If the controller's power is not on when replacing the battery, your date, time, and trend
data will be lost.
2
Remove the VVT Bypass II's cover.
3
Remove the battery from the controller, making note of the battery's polarity.
4
Insert the new battery, matching the battery's polarity with the polarity indicated on the controller's cover.
5
Replace the VVT Bypass II's cover.
6
Download the VVT Bypass II.

To clean the airflow sensor orifice

In less than .02% of installations, small particulates (micron/submicron-level) can build up in the airflow
sensor's internal orifice causing airflow restriction. Symptoms of airflow restriction are:
Diffuser noise increases over time
Space overcools at minimum airflow
Frequency of dampers opening to maintain flow setpoint increases over time
AHU fan speed increases over time
To check for these symptoms, set up trending of:
Damper position to look for long term increase
AHU fan speed for maintaining static pressure
If you see these symptoms, you should clean the flow sensor.
To clean the flow sensor
You can use a can of compressed air to clean the sensors while the controller is installed or on a repair
bench.
1
Disconnect the High and Low flow sensor tubes.
VVT Bypass II Controller
Installation and Start-up Guide
37
Troubleshooting
CARRIER CORPORATION ©2019
All rights reserved

Advertisement

Table of Contents
loading

Table of Contents