Quick Start Guide
2.4
Run Rosemount
When you first power up the Rosemount 6888 Xi, a short wizard program
guides you through the basic setup. Once configured, the Rosemount
6888Xi retains the setup, and the wizard will not repeat.
Procedure
1. Apply power to the Rosemount 6888 Xi.
Once boot-up is complete, the Quick Start Wizard screen appears.
With a dual channel Rosemount 6888 Xi, the wizard runs for both
channels in succession.
2. Press Enter to continue.
3. At the Sensor Type screen, use the Up and Down keys to select O
Do not select CO as this option is reserved for future use.
4. Press Enter to continue.
5. At the Device Type screen use the Up and Down keys to select
HART
6. At the Auto Cal Device screen, use the Up and Down keys to select
the calibration method to be used. The methods are defined as
follows:
7. Press Enter to continue.
Note
If you select SPS, IMPS, or Integral, you must still configure automatic
calibration as On. Verify other parameters, such as test gas values
and gas times, as well. Refer to the Rosemount 6888 Xi
Manual
8. When prompted by
to select Yes.
If you select No, the wizard restarts.
9. Press Enter to continue.
28
6888 Xi Quick Start Wizard
®
or FF (F
OUNDATION
None: Manual calibration with the standard probe housing
configuration.
SPS: Automatic calibration with the standard probe housing
configuration using the Rosemount SPS 4001B.
IMPS: Automatic calibration with the standard probe housing
configuration using the Rosemount IMPS.
Integral: Automatic calibration with the integral autocal probe
housing configuration.
for calibration setup details.
Setup Correct?
Fieldbus), whichever applies.
, use the Up and Down keys
October 2019
Reference
Emerson.com/Rosemount
.
2