Rosemount
probe to the Rosemount 6888 Xi. Use the following procedure to connect
the traditional architecture probe to the Rosemount 6888 Xi.
Note
The traditional architecture cable is provided at the specified length and is
ready for installation. The cable glands must be properly terminated to
maintain EMC/EMI noise protection.
Procedure
1. Run the 7-conductor cable between the traditional architecture
probe and the installation site for the Rosemount 6888 Xi.
Use new cable conduit or trough as needed.
2. Install the cable and lead wires to the probe per manufacturer's
instructions.
3. Install the cable at the probe housing and at the Rosemount 6888 Xi
enclosure.
Quick Start Guide
6888 Xi. A multi-conductor power/signal cable connects the
a) Unscrew locking nut from gland assembly and slide locking
nut back along cable.
b) Pull the gland body away from the plastic insert.
Use care not to damage the cable shield braid.
c) Insert the cable wires into the proper entry port in either the
probe housing or the Rosemount 6888 Xi enclosure.
d) At the probe housing, apply Teflon tape or similar sealing
compound to the tapered pipe threads. Thread the gland
body into the probe housing until properly seated.
e) At the Rosemount 6888 Xi enclosure, insert the gland body
into the left front cable port from the inside of the enclosure.
Use the rubber O-ring provided to seal the cable port.
f) Ensure the cable shield braid is evenly formed over the gray
insert.
When properly formed, the braid should be evenly spaced
around the circumference of the insert and not extend
beyond the narrow diameter portion.
g) Carefully press the gray insert into the gland body.
The grooves on the insert should align with similar grooves
inside the gland body. Press the insert in until it bottoms out
in the gland body.
h) Slide the locking nut up and thread it onto the gland body.
Tighten the locking nut so the rubber grommet inside the
Quick Start Guide
21