October 2019
JP2: Pins 2-3
JP5
Pins 1-2 internal power
Pins 2-3 external power
JP7
Pins 1-2 internal power
Pins 2-3 external power
JP8
Pins 1-2 internal power
Pins 2-3 external power
SW4 switch settings
Position 1: Off
Postion 2: Off
Position 3: Off
Position 4: Off
NOTICE
A. Except for JP5, JP7, and JP8 on IO board, jumper and switch settings are
factory set and are shown for reference only.
B. IO board 4-20 mA/HART loop power settings
JP5
JP7/JP8
Quick Start Guide
Pins 1-2 internal power Rosemount 6888 Xi to Rosemount 6888
Transmitter.
Pins 2-3 external power Rosemount 6888 Xi to Rosemount 6888
Transmitter (requires 2,500 resistor across J4, PR+ to PR-).
Pins 1-2 internal power Rosemount 6888Xi to DCS.
Pins 2-3 external power Rosemount 6888Xi to DCS.