Power consumption
Standby
• 220 V - 240 V / 50-60 Hz configuration
• 100 V - 120 V / 50-60 Hz configuration
During operation
• 220 V - 240 V / 50-60 Hz configuration
• 100 V - 120 V / 50-60 Hz configuration
Uninterruptible power supply (optional)
UPS Powerware 5115
UPS Powerware 5115
Environmental conditions
Room temperature
Maximum temperature change
Relative humidity
Magnetic field
Sunlight exposure
Environmental conditions (during storage)
In line with IEC721-3-1: 1K2 (CR 30-X) / 1K4 (CR 30-Xm) and 1M2
Temperature
Environmental conditions (during transport)
In line with IEC721-3-2: 2K2 and 2M2 with following restrictions:
Temperature
2386G EN 20120110
60 W
60 W
CR 30-X: max. 190 W
CR 30-Xm: max. 220 W
CR 30-X: max. 190 W
CR 30-Xm: max. 220 W
120 V
ABC ordering code: EGPSE
230 V
ABC ordering code: EGPTG
recommended: 20 °C to 25 °C
allowed: 15 °C to 30 °C
0.5 °C/min.
recommended: 30 % to 60 %
allowed: 15 % to 75 % (non-condensing)
compliant with EN 61000-4-8, Level 2
not to be operated in direct sunlight, max.
2500 lux
-25 °C to +55 °C
-25 °C to +60 °C
CR 30-X/CR 30-X
M
Equipment Information Sheet
55