If you notice conspicuous noise or smoke, disconnect the digitizer
immediately.
•
Do not pour water or any other liquid over the device.
•
The digitizer complies with the EN 60601-1 and UL 60601-1 standards for
Medical Electrical Equipment. This means that, although it is absolutely
safe, patients may not come in direct contact with the equipment.
Therefore the operator console must be placed outside a radius of 1.5 m
(EN) or 1,83 m (UL) around the patient (according to the local valid
regulation).
R = 1.5 m (1.83 m)
Patient environment
•
Perform no other operations on the digitizer than those described in this
document.
•
Switch the system off before moving it. When reaching the new position,
switch the system on again.
2386G EN 20120110
Patient environment
h= 2.5 m
(2.29 m)
CR 30-X/CR 30-X
M