AGFA CR 30-X User Manual page 62

Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

CR 30-X/CR 30-X
M
This device is intended for operation in the electromagnetic
environment given below. The user of the device should ensure
that it is used in such an environment.
Tests of Resistance
to Disruption
Conducted high
frequency disturbance
variables in
accordance with
IEC 61000-4-6
Radiated high
frequency disturbance
variables in
accordance with
IEC 61000-4-3
62
Remarks for HF-emission and immunity
IEC 60601
Test Level
Agreement
3 V
eff
3 V
150 kHz
to 80 MHz
3 V/m
3 V/m
80 MHz
to 2.5 GHz
Level of
Use portable and mobile radio sets
at a safe distance from the device
(including the leads) not closer
than the recommended protective
distance, which is calculated
according to the equation suitable
for the transmission frequency.
Recommended protective
distance:
d = 1.2
eff
d = 1.2
d = 2.3
P
With
transmitter in watts (W) in
accordance with the manufacturer
information on the transmitter
and d as the recommended
protective distance in metres (m).
The field strength of stationary
radio transmitters is lower than
the level of the agreement
frequencies in accordance with an
on-site investigation
Disruptions are possible near
devices that carry the following
symbol:
Electromagnetic
Environment
P
P
80 MHz to 800 MHz
P
800 MHz to 2.5 GHz
as the rated power of the
a
at all
b
.
2386G EN 20120110