M
Vibration
Environmental conditions for mobile installation (during transport)
In line with IEC721-3-5: 5K2 and 5M2 with following restrictions:
Vibration
Environmental conditions for mobile installation (during operation)
In line with IEC721-3-3: 3K2 and 3M1 with following restrictions:
Temperature
Relative Humidity
Vibration
Warming-up time
Cold start
Warm start
Physical emissions
Noise emission (sound power level according to ISO 7779)
• During scanning
• Standby
Heat emission
• Standby
• Average power consumption during
scanning
• Peak power consumption during scanning
End of Life
Estimated product life
(if regularly serviced and maintained
according to Agfa instructions)
56
Equipment Information Sheet
5-200 Hz (vertical, longitudinal, transversal
axis)
5-150 Hz (all axis), 1m/s², sinusoidal vibration
+15 °C to +30 °C
15% to 75% (non-condensing)
40-200 Hz; 1m/s²; sinusoidal vibration
fully operational after max. 30 min.
fully operational after self-test, on condition
that the digitizer:
• has not been switched off for more than
3 minutes.
• has been operational for at least 30 minutes.
max. 65 dB(A)
max. 55 dB(A)
1
60 W ≈ 204 BTU/h
CR 30-X: 85 W ≈ 290 BTU/h
CR 30-Xm: 103 W ≈ 351 BTU/h
CR 30-X: 190 W ≈ 648 BTU/h
CR 30Xm: 220 W ≈ 751 BTU/h
7 yrs.
1
1
1
1
2386G EN 20120110