M
Condition
Error during
writing on RF-tag
after scanning
process
Erasure lamp
failed during
scanning cycle
Image plate
detected in the
digitizer during
self-test
Digitizer cannot
detect the
cassette chip on
the image plate
tray e.g. tray is
missing or was
not inserted
correctly.
48
Operating CR 30-X/CR 30-Xm
Control PC message
"Error during writing on
cassette chip after
successful scan. Erase the
image plate again."
Digitizer errors
"Failure at erasure module.
Please restart digitizer. If
error remains, please
contact service. Image plate
must be erased after repair
again."
"Image plate during self-
test in the digitizer
detected. The image plate
was moved back into the
cassette. Please check
whether the image is
present at the PC and re-
scan the cassette, if
necessary. If the image is
present, the cassette was
not erased possibly. Erase
the cassette again in any
case."
"Cassette chip not readable
or cassette tray is missing.
Remove cassette."
Status
indicator
Confirm the message
with the OK button on
Red, flashing.
PC, remove the
After
cassette, push the
unlocking:
erase button on the
green,
digitizer and re-enter
constant
the cassette for a
manual erase cycle.
Restart the digitizer or
Red, constant
call service.
After power failure a
remaining image
plate, in the digitizer,
Red, flashing.
will be detected and
After
the forementioned
unlocking:
error message will be
green,
displayed.
constant
In order to release the
cassette, confirm the
message.
Red, flashing
Open the cassette and
check if the tray is
After
inserted correctly or
unlocking:
use another cassette
green,
and call service.
constant