AGFA CR 30-X User Manual page 49

Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

Condition
Communication
error, Ethernet
cable is not
plugged in
Time-out after 5
minutes if the ID
button of the
image processing
software was not
pressed.
Messages will stay active until the problem is solved or the dialog box message
on the control PC is confirmed by clicking the OK button.
Note: For error messages not listed in the table above, please refer to the
instructions within the displayed error message text.
2386G EN 20120110
Control PC message
"Image transmission failed.
System is retrying. If image
does not appear, please
restart the PC."
"ID button not pressed
within the timeout period.
Image plate was not
scanned. Cassette stays
clamped until ID data are
supplied by the operator."
Status
indicator
Check if Ethernet
cable is plugged in at
PC and digitizer.
Visually check if the
Red, flashing
Ethernet cable is
damaged. If the error
remains restart PC and
digitizer or call
service.
After ID
Confirm the message
button has
on the PC and press
been pressed:
the ID button of the
green,
image processing
flashing
software.
Operating CR 30-X/CR 30-Xm
CR 30-X/CR 30-X
M
Action