This device is intended for operation in the electromagnetic
environment given below. The user of the device should ensure
that it is used in such an environment.
Resistance to
Jamming Test
Discharge of static
electricity in
accordance with
IEC 61000-4-2
Fast transient
electrical
disturbance
variables / bursts in
accordance with
IEC 61000-4-4
Impulse voltages
(surges) in
accordance with
IEC 61000-4-5
Voltage
breakthroughs,
short term
interruptions and
variations in the
voltage supplied in
accordance with
IEC 61000-4-11
Magnetic field at
the supply
frequency (50/60
Hz) in accordance
with IEC 61000-4-8
• REMARK : U
is the alternating current in the network before the application of the test level.
r
2386G EN 20120110
IEC 60601
Test Level
+ 6 kV contact
discharge
+ 8 kV air
discharge
+ 2 kV for network
leads
+ 1 kV for entry
and outlet leads
+ 1 kV push-pull
voltage
+ 2 kV common
mode voltage
• < 5% U
(> 95%
r
breakthrough of
U
) for ½ period
r
• 40% U
(> 60%
r
breakthrough of
U
) for 5 periods
r
• 70% U
(30%
r
breakthrough of
U
) for 25
r
periods
• < 5% U
(95%
r
breakthrough of
U
) for 5 s
r
3 A/m
Level of
Agreement
Floors should consist of wood,
+ 6 kV contact
concrete or ceramic tiles. The
discharge
relative humidity must be at
+ 8 kV air
least 30%, if the floor is made of
discharge
synthetic material.
+ 2 kV for network
The quality of the voltage
leads
supplied should correspond to a
typical commercial or clinical
+ 1 kV for entry
environment.
and outlet leads
+ 1 kV push-pull
The quality of the voltage
voltage
supplied should correspond to
that of a typical commercial or
+ 2 kV common
clinical environment.
mode voltage
• < 5% U
(> 95%
r
breakthrough of
U
) for ½ period
The quality of the voltage supply
r
should correspond to that of a
• 40% U
(> 60%
r
typical commercial or clinical
breakthrough of
environment.
U
) for 5 periods
r
If the user wants the device to
• 70% U
(30%
r
work continuously, even when
breakthrough of
the energy supply is interrupted,
U
) for 25
it is recommended to use an
r
periods
energy supply free of
interruptions or a battery.
• < 5% U
(95%
r
breakthrough of
U
) for 5 s
r
Magnetic field at the network
frequency should correspond to
3 A/m
the typical values as they are in a
commercial and clinical
environment.
CR 30-X/CR 30-X
Electromagnetic
Environment Guidelines
Remarks for HF-emission and immunity
M
61