Troubleshooting 7
7.3
Spectrometer problems indicated by an error message in
OPUS
Error messages which refer to the spectrometer can be displayed in OPUS as follows:
•
in the Instrument Status dialog
•
by means of a yellow message bubble at the lower right end of the OPUS inter-
face
•
in a normal message window
i
In case of a spectrometer problem indicated by an error message, the OPUS status light
is yellow or red.
Error message in
OPUS
End of average life-
time is nearly
reached, spare part
will be required.
Temperature out of
range.
Laser is off or no laser
signal.
Scanner initialization
mode.
If an error message
appears which is not
listed above,.......
Table 7.3: Error messages in OPUS
128
Possible cause
The end of the specified life-
time of a spectrometer com-
ponent has nearly reached.
Temperature inside the spec-
trometer is too high.
Cause: Ambient temperature
is too high.
Defective laser
This error message appears
when you try to start a mea-
surement while the spec-
trometer is still initializing.
i
Further messages may be
displayed. As they do not
represent a spectrometer
problem you can ignore
them.
IFS 125M User Manual
Troubleshooting
Contact Bruker service.
i
Measuring is still possible.
Operate the spectrometer only in the
ambient temperature range specified
(18 - 35°C) by the manufacturer.
Contact Bruker service.
Wait until initialization has been fin-
ished.
...contact Bruker service.