The information in this manual may be altered without notice. BRUKER accepts no responsibility for actions taken as a result of use of this manual. BRUKER accepts no liability for any mis- takes contained in the manual, leading to coincidental damage, whether during installation or operation of the instrument.
Technical specifications ............. 19 Identification of parts ............20 Installation ................. 20 BCU05 HR to BCU05 HR-MAS ......23 How to change from BCU05HR to HR-MAS ....... 23 BCU05 HR-MAS to BCU05 HR ......25 How to change from BCU05HR-MAS to HR ....... 25 How to install a SENSE OF BEARING kit ......
Page 4
Contents 4 (41) BRUKER BCU05 Version 006...
Index W3M151933......................31 W4D125211 A..................... 33 W4M123713 A ....................35 W4M123717 A ....................30 W4S125315 A ..................... 34 W4S125337 A ..................... 32 W4S22941 B ....................... 29 W4S23389 A ....................... 28 BCU05 Version 006 BRUKER 5 (41)
BCU05 HR applications This unit contains ozone friendly coolant R404A. The unit may however only be discarted by authorized personnel. BCU05 Version 006 BRUKER 7 (41)
± 1°C under perfectly constant conditions of mains voltage, room temperature, air flow, etc. Operating principle The BCU05 is a cooling unit that reduces the temperature of an air input by typi- cally 65°C. Freon gas is circulating around an external heat exchanger. The system is com- pletely sealed.
Identification of parts Identification of parts see "BCU05 with BST Block diagram" on page 28 • Mains breaker with power control light P/N 75370 • Heat exchanger • Heat exchanger support stand P/N W1208509 Following parts are supplied : •...
BCU05 HR applications Installation See "BCU05 with BST Block diagram" on page 28 To install the system, proceed as follows : 1. Attach the support clamp (supplied with magnet) to the magnet. Insert the two half shells between clamp and exchanger.
Connect the air hose to the three way valve ex- haust. Figure 1.3. Glass ball joint (M151454B) 2. Take the heat exchanger support stand and position it between the magnet and the unit. Adjust the height. BCU05 Version 006 BRUKER 11 (41)
BCU05 HR applications Figure 1.4. Heat Exchanger support stand (M151454B) three way valve support stand heat exchanger main switch BCU 05 3. Position the cooling unit as far as possible from the magnet so as to minimise the effect of the magnet stray field.
If yes, decrease the air flow rate. Note: Please install the unit only on a hard floor. 6. If the BCU05 is controlled by a BVT3000/3300, connect the BVT plug to the connector called «BCU05» located on the BVT's front panel.
BCU05 HR applications Operating notes 1. The BCU05 is be powered by the BVT3000/3300 and is designed for 24 hours a day operation as long as the probe is not exchanged. If the probe is to be ex- changed then switch off the unit and wait for 10 minutes before handling. The exchanger tip must be de-iced before connecting on probe.
1. Remove the air vent grill and clean the air filter with compressed air. 2. Stop the BCU05 for a few hours. Switch off and disconnect the unit from probe. Blow dry air through the unit without cooling for ten minutes to remove the moisture from the heat exchanger.
BCU05 HR applications BCU05LT informations Technical specifications 1.6.1 This version of BCU05 with a longer heat exchanger (LT= 4m) is used with the largest magnets 750-800 MHz. It has the same technical specifications as the standard BCU05 unit. Identifications of parts 1.6.2...
BCU05 does not cool correctly The heat exchanger may be clogged with ice: • Switch off the BCU05 unit and wait for a period of 12 hours. • Disconnect the exchanger from the probe. • Put the exchanger on the floor.
Page 18
BCU05 HR applications 18 (41) BRUKER BCU05 Version 006...
± 1°C under perfectly constant conditions of mains voltage, room temperature, air flow, etc. Operating principle The BCU05 is a cooling unit that reduces the temperature of an air input by typi- cally 65°C. Freon gas is circulating around an external heat exchanger within a completely sealed system.
"Mounting the support clamp" on page 10 Installation See Installation BCU05 HR "Installation" on page 10 Put some vacuum grease on glass ball-joint before clamping the heat exchanger. Take care to keep the right distance between the probe glass ball-joint and the support clamp.
Installation Figure 2.2. Heat Exchanger support stand (W4M123713A) 2. Position the cooling unit as far as possible from the magnet so as to minimise the effect of the magnet stray field. BCU05 Version 006 BRUKER 21 (41)
3. Unscrew the HR trump output coupling 4. Screw the HR-MAS coupling 5. Install the pressure switch box 6. Connect the hoses and cables as shown on drawing "BCU05 HR-MAS Con- nection diagram" on page 33 and "BCU05 in operation" on page 35. BCU05 Version 006...
8. Screw the HR coupling 9. Install the pressure switch box 10. Connect the hoses and cables as shown on drawing "BCU in operation (W4M123717A)" on page 30 and "BCU05 with BST Block diagram" on page 28. BCU05 Version 006...
2. Install the kit as shown in picture below 3. Connect the hoses and cables as shown on drawing "BCU05 HR-MAS Con- nection diagram" on page 33 and drawing "BCU05 in operation" on page 4. Change the HR coupling with HR-MAS coupling (see chapter "How to change from BCU05HR to HR-MAS"...
BCU05 GAS FLOW BVT 3000 D N2 GAS OR DRIED AIR > - 50C SUPPORT CLAMP EXCHANGER BCU 05 MAINS GLASS BALL JOINT SUPPORT STAND B-CU 05 WITH BST BLOCK DIAGRAM ART NR : DRAWN APPROV MODIFICATION PRT NR : D.J.
SWITCH MAIN PLUG SOLID STATE RELAY CONNECTOR BROWN TO BVT 3000/BVT3300 WHITE BCU05 WIRING DIAGRAM ART NR : DRAWN APPROV MODIFICATION PRT NR : DWG NR : W4S22941B 10/92 06/95 DSK NR : W1K20345 BCU05 Version 006 BRUKER 29 (41)
BCU CN AIR IN BVT3X00 AIR IN MAIN CONNECTING HOSE MAS REMOTE CONTROL N2 GAS OR DRIED AIR > -50C HEATER BCU05 THERMOCOUPLE HR-MAS MAINS EXCHANGER SPIN RATE SUPPORT CLAMP GLASS BALL JOINT SUPPORT STAND ART: W1210342 DRAWN: VISA: DATE:...
Figure 5.7. BCU05 HR-MAS Wiring diagram 75617 HEATER THERMOCOUPLE PRESSURE SW THERMOCOUPLE COMPRESSOR HEATER TERMINAL BOX MAINS BREAKER 2A BLACK BLACK ON/OFF MAIN PLUG OPEN : IF DRIVEN BY BVT 3000 OR 3300 CLOSED : IN STAND ALONE MODE SWITCH...