Troubleshooting - Microlife WatchBP Office Instruction Manual

Professional automated office blood pressure monitor
Hide thumbs Also See for WatchBP Office:
Table of Contents

Advertisement

"Er 15"
Abnormal
result of
central blood
pressure
reading
"Er F"
The device
has gone into
"single fault
condition"
"Er A"
Flash memory
error
"HI"
Pulse or cuff
pressure too
high
"LO"
Pulse too low

Troubleshooting

Problem
Possible cause
No power (No
Power supply
is not properly
LCD display)
plugged in
Battery is fully
discharged
Cuff does not
Loose
connection of the
inflate properly
tube
Leakage of the
tube / bladder
No result
Device is in
displayed after
MANUAL Mode
measurements
The measuring signals are inaccurate so that no result can be
displayed. Read through the checklist for performing reliable
measurements and then repeat the measurement.
Single fault condition means that the measurement is aborted
to protect the patient from being harmed or the device
from being damaged. Re-position the cuff and repeat the
measurement. Replace the batteries if necessary. If the error
persists, contact Microlife or the local distributor.
Possible hardware fault. Try again. If the error persists,
contact Microlife or the local distributor.
The pressure in the cuff is too high (over 299 mmHg) OR the
pulse is too high (over 239 beats per minute). Relax for 5
minutes and repeat the measurement.
The pulse is too low (less than 30 beats per minute). Repeat
the measurement.
Solutions
Plug the power supply into the wall socket.
Recharge the rechargeable battery by plugging in the
power supply.
Make sure the tube of the cuff is securely connected
to the device.
Check for cracks on the tube or the bladder. Replace
the blood pressure cuff if necessary.
Switch to AUTO Mode and repeat the
measurements.
WatchBP Office
19
EN