9. Error Messages and Troubleshooting
When an error occurs during a measurement, the measurement is
interrupted and an error message displayed, e.g. „ERR 3".
Error
Designation
«ERR 1»
Signal too weak
Error signal
«ERR 2»
19
«ERR 3»
No pressure in
20
the cuff
30
Potential cause and remedy
The pulse signals on the cuff are too weak.
Re-position the cuff and repeat the measurement.
During the measurement interference signals were
detected on the cuff due to, for instance movement
or muscle tension. Repeat the measurement, not
moving your arm.
It is not possible to build up sufficient pressure in
the cuff. There may be a flaw in the cuff, causing
a pressure leak. Check that the cuff connection is
correct and that the cuff is not too loose around
your wrist. Replace the batteries if necessary.
Repeat the measurement.
Error
Designation
Potential cause and remedy
Abnormal result
The measuring signals are inaccurate and no result
«ERR 5»
can therefore be displayed. Read through the
checklist for performing reliable measurements
and repeat the measurement.
Pulse or cuff
The pressure in the cuff is too high (over 300
«HI»
pressure too high
mmHg) or your pulse rate is too high (over 200
beats per minute). Relax for 5 minutes and repeat
the measurement.*
«LO»
Pulse too low
The pulse rate is too low (less than 40 beats per
minute). Repeat the measurement.*
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you consider the measurement results to be unusual, please
observe the information in „Section 1.".