Bluetooth Connectivity; Rechargeable Battery And Power Adapter; Rechargeable Battery - Microlife WatchBP Office Instruction Manual

Professional automated office blood pressure monitor
Hide thumbs Also See for WatchBP Office:
Table of Contents

Advertisement

Bluetooth connectivity

Pairing the device
Press and hold the
seconds, untill the Bluetooth icon flashes and starts
pairing mode. The unique 6-digit device ID of the
unit is displayed. Connect the device and confirm
pairing. The Bluetooth icon is displayed on the
LCD screen of the device to show the presence of
Bluetooth connection.
Press and hold the Start/Stop button for 5 seconds to clear the
connection.

Rechargeable battery and power adapter

Rechargeable Battery

The device has a built-in, rechargeable Ni-
MH battery pack that can perform up to 400
measurement cycles on a full charge. The battery
can be recharged with the power adapter delivered
with the device. The empty battery indicator is
displayed when the battery is low.
• When using the device for the first time, charge
the battery until the recharge indicator turns to
green.
• The orange recharge indicator indicates that the
recharge is in progress.
• A green recharge indicator indicates that the
recharge is completed.
• A green and orange changing recharge indicator,
means that there is a charging error. Make sure
that of the correct Mains Adapter is used. If the
condition persists, contact Microlife or the local
distributor.
button for around 7
WatchBP Office
15
EN

Hide quick links: