Guidance and manufacturer's declaration – electromagnetic immunity
Immunity test
IEC 60601-1-2 test level
Electrostatic
±2kV, ±4kV, ±8kV, ±15kV air
discharge (ESD)
±8kV contact
EN 61000-4-2
Conducted
3V in 0,15 MHz to 80 MHz
disturbances
6V in ISM and amateur radio
inducted by RF
bands between 0,15 MHz and
fields
80 MHz
EN 61000-4-6
80% AM at 1 kHz
Radiated RF
Home Healthcare environment
electromagnetic
10 V/m
field
80 MHz to 2,7 GHz
EN 61000-4-3
80% AM at 1 kHz
Proximity fields
385 MHz - 27 V/m
from RF
450 MHz - 28 V/m
wireless
710, 745, 780 MHz - 9V/m
communications
810, 870, 930 MHz - 28 V/m
equipment
1720, 1845, 1970, 2450 MHz -
EN 61000-4-3
28 V/m
5240,5500, 5785 MHz - 9V/m
Electrical fast
±1kV SIP/SOP ports
transient/burst
±2kV AC port
EN 61000-4-4
100kHz repetition frequency
Power frequency
30A/m
Magnetic field
50 Hz or 60 Hz
EN 61000-4-8
Surge
±0,5kV ±1kV; ±2 kV, AC
Mains, Line to Ground
IEC 61000-4-5
±0,5kV ±1kV, AC Mains,
Line to Line
Compliance level
±2kV, ±4kV, ±8kV, ±15kV air
±8kV contact
3V in 0,15 MHz to 80 MHz
6V in ISM and amateur radio
bands between 0,15 MHz and
80 MHz
80% AM at 1 kHz
Home Healthcare
environment
10 V/m
80 MHz to 2,7 GHz
80% AM at 1 kHz
385 MHz - 27 V/m
450 MHz - 28 V/m
710, 745, 780 MHz - 9V/m
810, 870, 930 MHz - 28 V/m
1720, 1845, 1970, 2450 MHz
- 28 V/m
5240,5500, 5785 MHz - 9V/m
±1kV SIP/SOP ports
±2kV AC port
100kHz repetition frequency
30A/m
50 Hz
±0,5kV ±1kV; ±2 kV, AC
Mains, Line to Ground
±0,5kV ±1kV, AC Mains,
Line to Line
53
Electromagnetic
environment – guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material the
relative humidity level
should be at least 30%
Portable and mobile RF
communications equipment
should be used no closer to
any part of the product,
including cables, than 1.0m,
if the transmitter's output
power rating exceeds 1W
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site survey,
should be less than the
compliance level in each
b
frequency range
.
Interference may occur in
the vicinity of equipment
marked with this symbol:
Mains power supply should
be that of a typical
commercial or hospital
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.