Arjohuntleigh Maxi Air Instructions For Use Manual

Arjohuntleigh Maxi Air Instructions For Use Manual

Mattress
Hide thumbs Also See for Maxi Air:

Advertisement

Maxi Air
Instructions for Use
...with people in mind
001-35001-EN Rev 3 • 06/2017

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Maxi Air and is the answer not in the manual?

Questions and answers

Dawn
April 1, 2025

Are you supposed to clean the Maxi Air mattress when a patient soiled on the mattress with bleach wipes?

1 comments:
Mr. Anderson
May 14, 2025

The document does not mention using bleach wipes specifically. It states to follow the disinfection instructions provided and to use cleaning fluid (detergent and disinfectant), water, and cloths. Therefore, using bleach wipes is not confirmed.

This answer is automatically generated

Summary of Contents for Arjohuntleigh Maxi Air

  • Page 1 Maxi Air Instructions for Use ...with people in mind 001-35001-EN Rev 3 • 06/2017...
  • Page 2 It is mandatory to read the Instructions for Use. Design Policy and Copyright ® and ™ are trademarks belonging to the ArjoHuntleigh group of companies. © ArjoHuntleigh 2014. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice.
  • Page 3: Table Of Contents

    Product Specifi cations ......................5 Parts Designation ........................6 Product Description/Functions ..................... 7 Correct Placement of the Patient ..................8 Correct Placement of the Maxi Air air supply ..............9 Transfer ..........................10 Disinfection Instructions ..................... 14 Care and Preventive Maintenance ..................16 Troubleshooting ........................
  • Page 4 Intentionally left blank...
  • Page 5: Foreword

    Foreword Thank you for purchasing Manufacturer Information ArjoHuntleigh equipment. This product was manufactured by: Your Maxi Air™ is a part of a series of quality ArjoHuntleigh AB products designed for patient transfers, Hans Michelsensgatan 10 especially for hospital and professional 211 20 Malmö...
  • Page 6: Intended Use

    Preventive Maintenance” found in this IFU. The System The expected lifetime of the Maxi Air mattress, The Maxi Air (air supply and mattress) is an air is twenty (20) transfers. assisted lateral patient transfer system for use in lateral transfers and repositioning.
  • Page 7: Safety Instructions

    To avoid bodily injuries to the patient, use the Maxi Air in an environment where the temperature is below 32 °C (90 °F). Temperatures above 32 °C (90 °F) can cause the Maxi Air mattress surface to reach a temperature higher than 44 °C (111 °F). Safety Instructions...
  • Page 8: Preparations

    Patient” and “Product Specifi cations”. with section Parts Designation in this IFU. If any part is missing or damaged - do NOT Check that all parts of the Maxi Air is in place. use the product! Compare with the section Parts Designation in this IFU.
  • Page 9: Product Specifi Cations

    Product Specifi cations Product Specifi cations Maxi Air air supply Dimensions: 12,5 x 7 x 7”(317,5 x 178 x 178 mm) Weight: 11 lb (5 kg) Material: ABS Plastic Air Hose Lengths: 78” (1980 mm), 300” (7620 mm) Optional Power Cord Length: 180” (4570 mm) Maxi Air mattress Widths: L 34”...
  • Page 10: Parts Designation

    Parts Designation Parts Designation Maxi Air air supply Maxi Air mattress 1) Carrying handle 14) Connection points 2) Attachment hooks 15) Velcro lip 3) Socket inlet 16) Snap lock button 4) Push button (On/Off ) 17) Safety belts 5) LED mains power indicator...
  • Page 11: Product Description/Functions

    “Fig. 3”. Fig. 1 Operation and controls 1) Plug in the power cord from the Maxi Air air supply to an electrical outlet. 2) The LED will light up to indicate that the mains is turned on.
  • Page 12: Correct Placement Of The Patient

    Correct Placement of the Patient WARNING Always make sure the patient has the correct size of the Maxi Air mattress. If the Maxi Air mattress is too narrow, the Maxi Air mattress can tip over during transfer causing serious injuries...
  • Page 13: Correct Placement Of The Maxi Air Air Supply

    • Not in the way of the electrical outlet in use. • Not placed on the fl oor. An incorrectly placed Maxi Air air supply can cause hazards (e.g. interrupted infl ation). This can cause serious injuries to the patient and the caregivers.
  • Page 14: Transfer

    Position one caregiver on each side of the Fig. 10 patient, see “Fig. 10”. 3) Place the Maxi Air mattress with the labels upwards, underneath the patient. Follow local routines. 4) Make sure the patient is correctly placed on the Maxi Air mattress.
  • Page 15 8) Hang the Maxi Air air supply on the receiving surface. 9) Plug the power cord into an electrical outlet.
  • Page 16 To avoid bodily injuries, make sure the Maxi Air mattress is completely infl ated before starting the transfer. A partly infl ated Maxi Air mattress can cause the patient to hit the underlying surface, causing bodily injuries. The pull force will increase with a partial infl...
  • Page 17 Receiving side: make sure the Maxi Air mattress is centred, see “Fig. 17”. Fig. 16 14) Sending side: turn off the Maxi Air air supply and detach the hose from the Maxi Air mattress. Receiving side: fi rmly hold the handles until the Maxi Air mattress is completely defl...
  • Page 18: Disinfection Instructions

    fl uid. CAUTION 3) Roll up the air hose and lower it down into Do not lower down the Maxi Air air supply in cleaning fl uid. Scrub the hose with a brush. disinfectant solution. This could damage the 4) Rinse the air hose with water (approx.
  • Page 19 5) If cleaning fl uid cannot be removed in step 2.2 and step 2.4 spray water on the aff ected part and wipe off with disposable towels/ cloth. Repeat until all of the cleaning fl uid has been removed. Step 3 - Disinfection 1) Spray cleaning fl...
  • Page 20: Care And Preventive Maintenance

    Care and Preventive Maintenance Care and Preventive Maintenance The Maxi Air system is subject to wear and tear, and the following actions must be performed to make sure that the product remains within its original manufacturing specifi cation. WARNING To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow the recommended maintenance schedule.
  • Page 21 NOTE All Caregiver Obligations are to be checked when performing the Qualifi ed Personnel Service. Every Year : Maxi Air air supply has to be serviced according to the section “Care and Preventive Maintenance” schedule by qualifi ed personnel. Check and Change the Air Filter 1) Fold in the attachment hooks (A).
  • Page 22: Troubleshooting

    PROBLEM ACTION No air fl ow • Check power supply. • Check power cord connections on the Maxi Air air supply and the electrical outlet. Low pressure in the Maxi Air mattress. • Check air hose connections between the Maxi Air air supply and the Maxi Air mattress.
  • Page 23: Technical Specifi Cations

    Technical Specifi cations Technical Specifi cations Technical Data Weight 11 lb (5 kg) Expected life 5 years Protection class Not protected against ingress of water Environment pollution 2 maximum degree for use Operating force: Push button 30.0 N Maximum duty cycle For continuous use Sound Level 74 dBA MRI classifi...
  • Page 24 Allowed Combinations Mattress Mattress L MAS014000-WW Mattress XL MAS015000-WW ArjoHuntleigh Mattress XXL MAS016000-WW No other combinations are allowed. Air Supply    Air Supply 120 V MAS000001-US ArjoHuntleigh    Air Supply 230 V MAS000001-EU   ...
  • Page 25 European Union. Do not dispose these products as unsorted municipal waste. You can return your device and accessories to ArjoHuntleigh. You can also contact your local authorities for advice on disposal. The hospitals should follow local routine for disposal of contaminated Maxi Air mattress material.
  • Page 26: Labels On Maxi Air Mattress

    IEC 60601-1. The name and address of the manufacturer. Follow Instructions for Use Manufacturing date The shelf life (2 years) of the Maxi Air mattress LOT number starts from the manufacturing date on the LOT label on the outer package.
  • Page 27: Labels On Maxi Air Air Supply

    Push button MR unsafe Date Label 120V Maxi Air Model: 184445 MAS000001-US 120 V~60Hz 1100 W Class I Maxi Air Model: 184445 MAS000001-EU 230 V~50Hz 1200 W Date Label 230V Identification (inside) Corporate Logo Labels on Maxi Air air supply...
  • Page 28: Approvals And List Of Standards

    3) Turn the screws (A) to a locked (A2) position. Approvals and List of Standards Approvals and List of Standards The Maxi Air is compliant to IEC 60601-1 series, including applicable collateral standards and national deviations. Approvals and List of Standards...
  • Page 29: Electromagnetic Compatibility

    Electromagnetic Emissions - For all Equipment and Systems The Maxi Air is intended for use in the electromagnetic environment indicated below. The customer or the user of the Maxi Air should assure that it is used in such an environment. Emissions test...
  • Page 30 Electromagnetic Immunity - For all Equipment and Systems The Maxi Air is intended for use in electromagnetic environment specifi ed below. The customer or the user of the Maxi Air should assure that it is used in such an environment.
  • Page 31 Portable and mobile RF communications equipment should be used no closer to any part of the Maxi Air, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. 3 Vrms...
  • Page 32: Parts And Accessories

    6364346 - United Kingdom 6364347 - Europe 6364348 - South Africa 6364349 - India For disinfectant, contact your local representative. Air Filter 6305180 Handle Kit 700-32101 Maxi Air Mattress Maxi Air Cart MAS014000 MCA0003 MAS015000 (XL) MAS016000 (XXL) Parts and Accessories...
  • Page 33 Intentionally left blank...
  • Page 34 Intentionally left blank...
  • Page 35 AUSTRALIA FRANCE POLSKA ArjoHuntleigh Pty Ltd ArjoHuntleigh SAS ArjoHuntleigh Polska Sp. z o.o. 78, Forsyth Street 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 O’Connor CS 70133 PL-62-052 KOMORNIKI (Pozna ) AU-6163 Western Australia FR-59436 RONCQ CEDEX Tel: +48 61 662 15 50 Tel: +61 89337 4111 Tél: +33 (0) 3 20 28 13 13...
  • Page 36 Hans Michelsensgatan 10 211 20 Malmö, Sweden www.arjohuntleigh.com ArjoHuntleigh is a world-leading provider of integrated products and solutions that improve the lives of patients and residents with reduced mobility. We help healthcare facilities deliver wellness and eff ective everyday care, early mobilisation, safe patient handling, venous thromboembolism prevention, pressure injury prevention, hygiene routines, bariatric care and diagnostics.

Table of Contents