Arjohuntleigh Malibu Instructions For Use Manual page 34

Hide thumbs Also See for Malibu:
Table of Contents

Advertisement

Fig. 1
1
2
Fig. 2
Fig. 3
Fig. 4
A
Clean and Disinfect with Disinfection Unit –
with or without Air Spa System (40 Steps)
Remove visual residue (Step 1-7 of 40)
1 Raise bathtub to convenient working position.
2 Make sure all functions are turned off.
3 Open drain to empty any left over water.
4 Detach and place the loose parts such as arm rests, seat of
the transfer chair, the neck pillow and other accessories such
as safety belts on the bottom of the bathtub. (See Fig. 1)
5 Make sure the transfer chair is in the bathtub and not
outside.
6 Shower and rinse the bathtub, the lift arm and the transfer
chair with water to remove visible residue. Start from the
top and move downwards.
7 Shower and rinse the loose parts and the accessories in the
bottom of the bathtub with water to remove visible residue.
Turn accessories and loose parts and rinse again.
Clean (Step 8-24 of 40)
8
Put on protective gloves and eye wear. (See Fig. 2)
9 Open the disinfection cover and lift out the disinfection
spray handle. (See Fig. 3)
10 Check the level of disinfectant.
11 Push down the disinfection button (A) on the control panel
for 5 s. The LED will light up green (See Fig. 4)
12 If the bathtub has an Air Spa system:
• The automatic disinfection of the Air Spa system will
start. Approximate time for the automatic disinfection is
3-4 minutes. In about 30 s after the start of the automatic
disinfection the disinfection spray handle is activated and
can be used.
13 Direct the disinfection spray handle against the bottom of
the bathtub and press the trigger.
34
Continue with the steps on the next page.

Hide quick links: