Download Print this page
Arjohuntleigh Malibu Quick Reference Manual

Arjohuntleigh Malibu Quick Reference Manual

Medical bathtub
Hide thumbs Also See for Malibu:

Advertisement

Quick Links

QUICK REFERENCE GUIDE - CLEANING AND DISINFECTION
Malibu/Sovereign
Disinfect with disinfection unit
Remove visual residue
1
Put on protective gloves
and eye wear.
Clean
5
Lift out the disinfection spray
handle.
8
Spray disinfectant into the bathtub until
the flow meter (A) shows correct mixing
value according to the Instructions for
use. Adjust if needed (B).
Mixing temperature 20°C (68°F).
10
Spray disinfectant
on all internal
bathtub surfaces, lift
arm, transfer chair,
the loose parts and
on all sides of the
accessories.
13
Open the drain.
Open the drain.
Rinse all parts with
Rinse all parts with
plenty of water
plenty of water
having a temperature
having a temperature
of 25 °C (77 °F).
of 25 °C (77 °F).
Disinfect
16
Spray disinfectant in the drain. Close
the drain.
ArjoHuntleigh AB
Hans Michelsensgatan 10
211 20 Malmö, Sweden
www.arjo.com
2
3
Turn off all
functions.
Make sure the
tran fer chair
is inside the
Open the drain. Detach and place all
bathtub.
accessories at the bottom of the bathtub.
6
7
If the bathtub has an Air Spa system:
Push down the
The automatic disinfection will start. Approximate time
disinfection
for the automatic disinfection is 3–4 minutes.
button
30 seconds after the start of the automatic disinfection
for 5 seconds.
the disinfection spray handle can be used.
11
Spray into the surface of the overflow
outlet.
14
Soak a cloth in disinfectant. Wipe all other contact
areas such as the outside of the bathtub, hand
control and the wheel chassis with the cloth.
17
Spray disinfectant
on all internal
bathtub surfaces, lift
arm, transfer chair,
the loose parts and
on all sides of the
accessories.
08.AZ.00_6EN • 09/2020 • Page 1/3
B
A
12
18
Soak a cloth in disinfectant. Wipe all other contact
areas such as the outside of the bathtub, hand control
and the wheel chassis with the cloth.
4
With water, shower and
rinse the bathtub, lift arm,
transfer chair, the loose
parts and all sides of the
accessories.
9
Spray disinfectant in the drain. Scrub the
drain with a brush. Close the drain.
Scrub all internal bathtub surfaces, lift arm,
transfer chair, the loose parts and the accessories
thoroughly.
15
Wet a new cloth with water and remove the
disinfectant. Occasionally rinse the cloth in
running water.
Continue with the steps on the next page

Advertisement

loading

Summary of Contents for Arjohuntleigh Malibu

  • Page 1 QUICK REFERENCE GUIDE - CLEANING AND DISINFECTION Malibu/Sovereign Disinfect with disinfection unit Remove visual residue Turn off all With water, shower and functions. rinse the bathtub, lift arm, Make sure the transfer chair, the loose tran fer chair parts and all sides of the...
  • Page 2 Disinfect with disinfection unit (continued) If the bathtub has an Air Spa system: Push the disinfection button and it will shorten the sequence Allow appropriate for the automatic disinfection cycle. Reduced time is up to contact time Push down the 35 seconds.
  • Page 3 Attach the neck pillow, remove disinfectant. Occasionally safety belt and rest of rinse the cloth in running water. Leave to air dry. the accessories. ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö, Sweden www.arjo.com 08.AZ.00_6EN • 09/2020 • Page 3/3...

This manual is also suitable for: