4 Start up and shut down
4.3 Check the pressure
Compressed air connection on devices.
Contaminated and humid compressed air leads to premature wear and
tear.
▶ Supply dry, clean and uncontaminated compressed air according to
ISO 7494-2 only.
▶ Insert the test gauge (Mat. no. 04118731) between the MULTIflex
(LUX) Kupplung / MULTIflex LED Kupplung couplings and the medi‐
cal device, and check the following pressures:
ð - Drive air: 2.1 to 3.5 bar (30 to 51 psi) (recommended: 2.8 bar (41
psi))