Start Up And Shut Down; Mounting The Sirona Coupling; Check The Pressure - KaVo PROPHYflex 3 Instructions For Use Manual

2018 for sirona connection
Hide thumbs Also See for PROPHYflex 3:
Table of Contents

Advertisement

Instructions for use PROPHYflex 3 - 2018 for Sirona connection - 1.006.9920, - 1.006.9927

4 Start up and shut down

Hazard from non-sterile products.
Infection hazard for dentist and patient.
▶ Before first use and after each use, prepare and sterilise the medical device and
accessories appropriately.
Dispose of the product in appropriate manner.
Infection hazard.
▶ Before disposal, reprocess and sterilise the product and accessories appropri-
ately.
Damage from soiled and moist cooling air.
Contaminated and moist cooling air can cause malfunctions.
▶ Make sure that the supply of cooling air is dry, clean, and uncontaminated accord-
ing to EN ISO 7494-2.

4.1 Mounting the Sirona coupling

Detachment of the medical device during treatment.
A medical device that is not properly locked can detach from the Sirona coupling dur-
ing treatment.
▶ Before each treatment, check if the medical device is securely locked onto the
Sirona coupling.
▶ Screw the Sirona coupling onto the turbine hose.
The amount of water in the spray can be regulated by turning the spray ring at the
Sirona Regulatable coupling.
The water volume has a crucial effect on the cleaning efficiency and on dust develop-
ment.
▪ Small water volume = little cleaning efficiency and a lot of dust.
▪ Large water volume = great cleaning efficiency and less dust.
Note
As the PROPHYflex 3 - 2018 SIRONA does not require any light or airspray, these
functions will need to be deleted.

4.2 Check the pressure

Contaminated or moist compressed air at the compressed air connection
Premature wear
▶ Ensure that the cooling air is dry, clean and uncontaminated in accordance with
EN ISO 7494-2.
A drive pressure of 3.2 bar (46 psi) is required for operating the PROPHYflex.
▶ Install the Sirona test manometer test star between the hose and coupling.
4 Start up and shut down | 4.1 Mounting the Sirona coupling
WARNING
WARNING
NOTICE
WARNING
NOTICE
11 / 26