Hologic ThinPrep Operator's Manual page 62

Integrated imager
Hide thumbs Also See for ThinPrep:
Table of Contents

Advertisement

The Microscope - a customized microscope with imaging camera, slide ID camera, automated
stage, stage controls and touch screen interface.
The trinocular head - a tilting binocular observation tube and a fixed, straight tube for the imaging
camera. The light path and camera focus have been optimized by placement of spacers in the
assembly of the optical components. Do not add or remove spacers or risers.
If an optional telescoping head is being used, be sure to use the specific riser that Hologic
supplies.
One ocular has a diopter adjustment ring to provide common focus capability.
If the optional riser is being used, do not use it in conjunction with the telescoping head. Use one
or the other, but not both.
CAUTION: Only use Hologic-supplied eyepieces and objective lenses. DO NOT substitute eye-
pieces or objectives.
Eyepieces - 10X magnification with a field size of 22 mm.
Objectives - 4X, 10X and 40X objectives are mounted on the revolving nosepiece at production.
They are specifically compatible with the supplied eyepieces and the camera for the imaging
system. They should not be substituted with any other objectives.
The other object in the nosepiece is the magnetic 10X position sensor. It must not be removed.
An optional 20X objective is available. (Refer to Chapter 8, Ordering Information.) It can be
installed by the operator. If the 20X objective is installed, the objectives should be positioned as
shown in Figure 2-4.
Standard Configuration for
Magnetic 10X
MAGNET
position sensor.
Do not move.
40X
User interface touch screen and mounting rail - the touch screen height can be adjusted by moving
the screen up or down along the mounting rail. The tilt and rotational angle of the screen may be
adjusted by loosening the adjustment knobs, changing the tilt and rotation, and then tightening
each knob.
4X, 10X and 40X
Open
4X
10X
Front
Figure 2-4 Positions of Objectives in the Nosepiece
I
NSTALLATION
Optional Configuration for
4X, 10X, 20X and 40X
MAGNET
4X
®
ThinPrep
Integrated Imager Operator's Manual
2
40X
20X
10X
Front
2.5