Table 16. Summarized Results of Within-instrument Study
Dx
Imager
TIS
NILM
I2
TIS
ASCUS
I2
TIS
LSIL
I2
TIS
ASC-H
I2
TIS
AGUS
I2
TIS
HSIL
I2
TIS
CANCER
I2
TIS
UNSAT
I2
Between-instrument Reproducibility
Between-instrument reproducibility results were derived from the clinical study. In the clinical
study, three (3) cytotechnologist/pathologist pairs reviewed slides on different instruments.
In Table 17, the between-instrument results are summarized for each diagnostic category of
slides (according to adjudicated truth results). For each grouping, the following metrics are
reported:
% Abnormal
The proportion of slides for which any abnormal diagnosis was recorded.
(For NILM or UNSAT slides, the % Normal column is used to record the proportion that are
not abnormal).
% Category+
The proportion of slides for which the site diagnosis was equal to or higher than the slide's
adjudicated category.
 
%
%
Abnormal
Category+
75.9%
75.9%
71.9%
71.9%
97.3%
93.2%
96.0%
94.0%
93.3%
86.7%
100%
83.3%
63.0%
51.9%
55.6%
48.1%
98.0%
77.3%
97.3%
71.3%
100%
46.7%
100%
53.3%
%
%
Normal
N/A
% zero
69.6%
11.0%
70.4%
78.1%
4.3%
78.4%
13.3%
25.0%
5.0%
28.1%
3.3%
2.8%
0.7%
4.0%
0.0%
6.7%
0.0%
0.0%
6.7%
35.7%
10.0%
44.4%
0.0%
2.0%
0.7%
2.7%
0.0%
0.0%
0.0%
0.0%
72.2%
40.0%
72.2%
85.7%
36.7%
94.7%
MAN-05359-001 -001 Rev. 001 page 15 of 32
Abnormal FOV
Median
0
0
6
7
14
15
11.5
14
2
2
20
20
22
22
0
0

Hide quick links: