Hologic ThinPrep Operator's Manual page 125

Integrated imager
Hide thumbs Also See for ThinPrep:
Table of Contents

Advertisement

4
O
PERATION
The Integrated Imager images the cell spot.
To ensure that the focus and illumination requirements for imaging are met, the system
Note:
disables manual control of the X, Y axis stage control knobs, focus knobs and light adjust-
ment. Manual control is returned to the operator after the imaging process is complete.
During imaging a green progress bar represents
how much of the cell spot has been imaged.
Do not remove the slide from the stage during imaging. To cancel imaging, press the Cancel button.
4.10
®
ThinPrep
Integrated Imager Operator's Manual
When the cell spot has been imaged, the Inte-
grated Imager performs functional checks prior to
completion.
Figure 4-10 Imaging Slide in Process
Figure 4-11 Imaging Complete