Storage; Maintenance And Service; Measures For The Product Disposal - Primes CompactPowerMonitor Series Operating Manual

Laserdiagnosticssoftware lds, powermonitorsoftware pms
Table of Contents

Advertisement

14

Storage

Please note before storing:
NOTICE
Damage/destruction of the device caused by leaking or freezing cooling water
Leaking cooling water can damage the device. Storing the device at temperatures near or
below freezing and without emptying the cooling circuit completely can damage the device.
X Empty the lines of the cooling circuit completely.
X Even when the lines of the cooling circuit have been emptied, a small amount of residual
water will remain in the device at all times. This may leak out and end up inside the device.
Close the connector plug of the cooling circuit with the included sealing plugs.
X Store the device in the original PRIMES transport box.
NOTICE
Damaging/Destruction of the flow rate meter
The flow rate meter is not designated for high rotational speed.
X Do not use compressed air for emptying the cooling circuit.
15

Maintenance and service

The operator is responsible for determining the maintenance intervals for the measuring device.
PRIMES recommends a maintenance interval of 12 months for inspection and validation or calibration.
If the device is used only sporadically, the maintenance interval can also be extended up to 24 months.
16

Measures for the product disposal

Due to the Electrical and Electronic Equipment Act ("Elektro-G") PRIMES is obliged to dispose PRIMES mea-
suring devices manufactured after August, 2005, free of charge. PRIMES is a registered manufacturer in the
German "Used Appliances Register" (Elektro-Altgeräte-Register "EAR") with the number
WEEE-Reg.-Nr. DE65549202.
Provided that you are located in the EU, you are welcome to send your PRIMES devices to the following ad-
dress where they will be disposed free of charge (this service does not include shipping costs):
PRIMES GmbH
Max-Planck-Str. 2
64319 Pfungstadt
Germany