CompactPowerMonitor CPM
NOTICE
Damage/Destruction of the device due to overpressure
X The maximum permissible water inlet pressure must not exceed 4 bar.
8.3

Humidity

The device must not be operated in a condensing atmosphere. The humidity has to be considered in order
to prevent condensates within and outside the device.
The temperature of the cooling water must not be lower than the dew point (see Tab. 8.2 on page 21).
NOTICE
Damage/Destruction of the device due to condensing water
Condensation water inside of the device can lead to damage.
X Mind the dew-point in Tab. 8.2 on page 21.
Do only cool the device during the measuring operation. We recommend starting the cooling approx. 2 min-
utes before the measurement and terminating it approx. 1 minute after the measurement.
Tab. 8.2:
Dew point Diagram
Example
Air temperature:
Relative humidity:
The cooling water temperature cannot fall below 14 °C.
Revision 02/2019 EN
40
35
30
25
20
15
10
5
0
0
5
10
15
Air temperature in °C
22 °C
60 %
20
25
30
35
40
100
80
70
60
50
40
30
20
10
21