CompactPowerMonitor CPM
9.3
Connect the safety interlock
The device can be damaged when the water flow rate is too low, the inlet temperature T
temperature difference T
D
laser in this case.
If the water flow rate is too low, the inlet temperature T
great, then Pin 1 and Pin 4 are connected. Pin 1 and Pin 3 are connected when required by the operating
condition values.
NOTICE
Damaging/Destruction of the device
If the safety interlock is not connected, this may lead to damages to the device due to over-
heating.
X Make sure to connect the safety interlock of the laser control in a way that ensures that the
laser is turned off whenever this connection is interrupted.
A suitable connection cable with a device plug and free ends is included in the scope of delivery
Pin assignment socket
(top view, plug-in side)
4
3
1
Tab. 9.2:
Pin assignment safety interlock
Revision 02/2019 EN
too great. The safety interlock protects the device from damages by turning off the
Pin
Wire color
Function
1
Brown
Mutual pin
3
Blue
Connected with Pin 1 when ready for operation
4
Black
Connected with Pin 1 when in safety interlock mode
(water flow rate too low)
is too high, or the temperature difference T
in
too high, or the
in
is too
D
25