7.4

Install the CompactPowerMonitor CPM

DANGER
Serious eye or skin injury due to laser radiation
If the device is moved from its calibrated position, increased scattered or directed reflection
of the laser beam occurs during measuring operation (laser class 4).
X Mount the device so that it cannot be moved by an unintended push or a pull on the cables
or hoses.
NOTICE
Damage/Destruction of the device
Too long fixing screws can damage internal components in the device.
X The fastening screws must not be screwed in more than 8 mm into the housing.
There are fastening screw threads M4 and M6 in the bottom of the housing which are intended for the
mounting on a customer's mount. We recommend screws of the strength class 8.8 and a tightening torque
of 5 N∙m.
Fig. 7.2:
Fastening screw threads in the bottom of the housing of the CPM C-9