Analog Output - Primes CompactPowerMonitor Series Operating Manual

Laserdiagnosticssoftware lds, powermonitorsoftware pms
Table of Contents

Advertisement

CompactPowerMonitor CPM
9.7

Analog output

The CompactPowerMonitor CPM has an analog voltage output (analog output) that emits a voltage value
analogous to the measured laser power. The analog signal is effected via the 4-pin device socket M8 (see
Fig. 5.1 on page 12). The output voltage amounts to a maximum of 10 V. The output voltage of 10 V is
scaled to the maximum output value of the connected device (see Tab. 9.3 on page 31). A suitable con-
nection cable is included in the scope of delivery.
The load resistance at the analog output should not be smaller than 100 kOhm.
An output voltage of 1 V
equals approx.
Tab. 9.3:
Output voltage and laser power
When power is supplied through the USB port, it is not necessary to hook up 24 V through Pin 1 and Pin 2.
Pin assignment socket
(top view, plug-in side)
Tab. 9.4:
Socket assignment of the analog output
Revision 02/2019 EN
CPM C-9
CPM F-1
1 000 W
250 W
Pin
Wire color
Function
1
Brown
24 V (input power supply)
2
White
Ground for the power supply
3
Blue
Ground for the analog signal
4
Black
Analog signal 0 – 10 V (output)
CPM F-10
CPM F-20
1 000 W
2 500 W
CPM F-30
4 000 W
31