15 ISE troubleshooting
List of ISE flags
Air Isecal
Meaning
Message ID
Possible cause
Recommended actions
? Cal
Meaning
Message ID
Possible cause
Recommended actions
Cal Error
Meaning
Message ID
Possible cause
Recommended actions
Roche Diagnostics
E-72
o
The length of the calibrator segment was too short.
o
The end of the calibrator segment was not detected. It takes too long for a
calibrator segment to reach the electrodes.
91
Priority
5
o
Air bubbles are present.
o
The ISE tubing is leaking or blocked.
o
Ensure that the ISE Calibrator Indirect on the ISE unit is available and check the
level of fluid; replace if necessary.
o
Repeat the measurement.
o
Check the ISE tubing for leaks and blockages.
A result where its calibration is flagged.
110
Priority
29
The calibration was in question and has been flagged, however the calibration could
be used to calculate the result (compare with Cal Error where the calibration did not
provide a usable result).
Check and redo calibration if needed.
No valid calibration data available.
43
Priority
24
Caused by an alarm that occurred during calibration because the calibration could
not provide a usable result.
Check the flags of the calibration and proceed to deal with the calibration first.