5 Daily operation
Preparing the system
Step
5
Prepare the reagents.
6
Prepare the cuvettes.
7
Perform reagent mixing.
8
Perform calibrations.
Table B-2
Steps for preparing the system
Roche Diagnostics
B-14
User action
Deal with at least all red and yellow buttons.
1. Press a reagent button to check the details.
Replace the empty reagents.
2. Press
.
3. Open the main cover
4. Remove the bottle.
5. Press
to confirm.
6. Remove the remaining bottles of the set.
7. Press
.
8. Place the new bottle.
9. Press
to confirm.
10. Load the remaining bottles of the set.
11. Press
to proceed to the next stage in the
Prepare wizard.
Replace all red cuvette segments.
1. Press the cuvette button.
2. Open the main cover.
3. Replace the cuvette segment.
4. Press
to confirm the replacement.
5. Replace the remaining segments that need
replacing.
6. Press
to proceed to the next stage in the
Prepare wizard.
1. Select the test(s).
2. Press
.
Perform the calibrations.
1. Check the test selection.
2. Press
.
3. Place the calibrators.
4. On the placement screen, press
5. Press
.
6. Validate the results.
Operator's Manual · Version 3.1