Importing Data; Importing Applications - Roche Cedex Bio System Operator's Manual

Hide thumbs Also See for Cedex Bio System:
Table of Contents

Advertisement

Cedex Bio System

Importing data

Importing applications

Roche Diagnostics
Operator's Manual · Version 3.1
f
Utilities > Import
You can import the following kinds of data:
o
Application data (application definitions)
o
Software (software updates)
o
Database (as exported using Utilities > Export > Database)
o
Certificates (digital records that ensure the authenticity of reagent barcodes)
o
Extra wash cycle and mixing rule definitions
Importing an application consists of two steps:
1. Importing the data to the instrument by reading a barcode or a data file. This step
stores the data on the instrument.
2. Installing the application. This step activates the application on the instrument
and so makes it available for use.
a To import application data
1 Do one of the following:
o
Choose Utilities > Applications > Laboratory Parameters, then continue
with step 2.
o
Choose Utilities > Import > Application, then continue with step 4.
2 Press
.
3 Press Import Application.
A screen is displayed, asking you to scan the barcode or to import the data from
the USB stick.
4 Do one of the following:
If...
You intend to scan the barcode:
You intend to import from the USB stick:
The system checks whether there is already an application on the system with
identical application code and short name.
6 Special operations
Do this...
1. Scan the barcode from the barcode
sheet.
1. Press
.
2. Insert the USB stick.
3. Press
.
4. Select the directory that contains the
application file.
You recognize the application pack-
ages by their file extension .tsb.
5. Press
to confirm the selection.
Importing data
B-111