4 Define the lot values. Do one of the following:
If...
Do this...
The lot data are
1. Scan the barcode.
available from a bar-
code
You want to type the
1. Press
data manually
2. Type the name (up to 10 alphanumeric characters) and press
3. Type the material code and press
4. Type the lot ID (up to nine alphanumeric characters) and press
5. Type the expiration date and press
6. Type the number of calibrators (cups) you need to place on the
7. Press
8. Press *Add new test*.
9. Press
10. Press one of the test buttons.
11. Press
12. Define the first calibration value (target value).
13. Define the next calibration value.
14. Select the value for Last is Water.
15. Press
Calibration values (target values) in method sheets may be defined in increasing order of
concentration. Always define the values in decreasing order on the Cedex Bio instrument.
The screen is displayed again that lists all installed calibrator lots.
The new lot is selected.
.
Refer to the package insert.
Use the date format as indicated on the screen.
instrument. Use the information given on the package insert.
(Exclude calibrators that use system water from this number. See
step 14.)
to confirm the definitions.
.
A screen is displayed that shows all tests that have this calibrator
defined for them and that are not yet assigned to this lot.
.
You can define up to six calibration values, and you must define
them in descending order.
Press On if you want to calibrate with system water as zero cali-
brator. (In this case, no cup needs to be placed on the system. (See
step 7).
Press Off if you want to use a special zero calibrator for the cali-
bration. (The special zero calibrator needs to be placed on the
sample area.)
to confirm the definitions.