Cedex Bio System
List of ISE flags
>
Meaning
Message ID
Possible cause
Recommended actions
Air Fluid
Meaning
Message ID
Possible cause
Recommended actions
Roche Diagnostics
Operator's Manual · Version 3.1
This section lists the general and ISE-specific flags in alphabetic order and provides
information on each of the flags.
e
For information on flags that concern absorbance measurements only, see List of flags on
page D-28.
Where a flag description gives a list of recommended actions, work through each step in
turn until you have solved the problem. Proceed to the next step only after you have com-
pleted the previous one.
Result generated by re-performing the test with the same running parameters
(repeated).
120
Priority
32
The user initiated the repeat.
No action required.
o
The length of the sample segment was too short.
o
The end of the sample segment was not detected. It takes too long for a sample
segment to reach the electrodes.
19
Priority
4
o
Air bubbles are present.
o
The ISE tower is clogged.
o
The dosing from the sample probes is inaccurate.
o
The ISE tubing is leaking, blocked, or not conditioned.
o
Mixing is too vigorous.
o
Check the quality of the sample.
o
Check the sample tubes.
o
Check the ISE tower and ISE tubing for leaks and blockages.
o
Repeat the measurement.
15 ISE troubleshooting
List of ISE flags
E-71